AirSense™ 10 CPAP Machine


The AirSense 10 AutoSet is a premium auto-adjusting pressure therapy device that automatically adjusts pressure levels to fit your changing needs and to ensure you’re receiving the lowest pressure necessary. The AirSense 10 AutoSet is also set to deliver therapy at the most comfortable temperature and humidity levels when used with ResMed’s new HumidAir™ heated humidifier and ClimateLineAir™ heated tube.

Engineered to be intuitive and easy-to-use, you can simply plug in the AirSense 10 AutoSet device and press Start

 

TECHNICAL SPECIFICATIONS 

90W power supply unit 
AC input range: 100–240V, 50–60Hz 1.0–1.5A, Class II 
115V, 400Hz 1.5A, Class II (nominal for aircraft use) 
DC output: 24V 3.75A 
Typical power consumption: 53W (57VA) 
Peak power consumption: 104W (108VA) 
Sensors 
Pressure sensor: 
Internally located at device outlet, analog gauge pressure type, -5 to +45 cm H2O 
Flow sensor: 
Internally located at device inlet, digital mass flow type, -70 to +180 L/min
Maximum single fault steady pressure 
Device will shut down in the presence of a single fault if the steady state pressure exceeds: 
30 cm H2O for more than 6 sec or 40 cm H2O for more than 1 sec. 
Sound 
Pressure level measured according to ISO 17510-1:2007 (CPAP mode): 
SlimLine: 
26.6 dBA with uncertainty of 2 dBA 
Standard: 

Physical - device and water tub 
Dimensions (H x W x D):
4.57" x 10.04" x 5.91" (116 mm x 255 mm x 150 mm) 
Air outlet (complies with ISO 5356-1:2004): 22 mm 
Weight (device and standard water tub): 
44 oz (1248 g) 
Temperature 
Maximum heater plate: 
154°F (68°C) 
Cut-out: 
165°F (74°C) 
Maximum gas temperature: 
 106°F ( 41°C) 
Air filter 
Standard: 
Material: Polyester non woven fiber 
Average arrestance: >75% for ~7 micron dust 
Hypoallergenic:
Material: Acrylic and polypropylene fibers in a polypropylene carrier Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron dust 

Aircraft use 

ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel. 

Wireless module 

Technology used:
CDMA (USA and Canada only) 2G GSM (all regions except USA and Canada) 
FCC ID: 2ACHL-AIR10CD 

The AirSense 10 device complies with FCC Rules. 
The AirSense 10 device should be used at a minimum distance of 0.8" (2 cm) from the body during operation. Additional information regarding the FCC Rules for this device can be found on www.resmed.com. 

Operating pressure range 
AutoSet, AutoSet For Her, Elite, CPAP 4 to 20 cm H2O 

Supplemental oxygen 
Maximum flow: 4 L/min 
Pneumatic flow path